Ph meter reader is use for checkmating ,e.g those into skincare for you to be sure your cream,oil,soap are active and safe you need ph meter to do that.
easy to use: simply immerse the ph meter electrode in the solution to be tested, readings are quickly display on a large lcd display.
wide application: ideal instrument for any aquarium, the fishing industry, swimming pools, spa, school laboratories, food & beverage, drinking water etc. fast and accurate results: no need to work with those annoying color charts and test strips, its also comes with Protective fluid.
compact portable design: pocket fitting ph meter that’s also lightweight and compact.
HOW TO CALIBRATE
this is the SY-1030 / pH-801/902 style pen pH meter (very common). It has the MODE/CAL and HOLD/TEMP buttons and your buffer powders (4.01, 6.86, 9.18), so here’s the exact way to calibrate THIS type.
Step 1: Prepare the buffer solutions
Each sachet is for 250 ml.
• pH 6.86 → dissolve in 250 ml distilled water
• pH 4.01 → dissolve in 250 ml distilled water
• pH 9.18 → optional (for soap / alkaline products)
Use clean cups. Label them if possible.
⸻
Step 2: Power on & rinse
• Remove the cap
• Rinse probe with distilled water
• Gently shake or blot dry (don’t rub)
⸻
Step 3: Calibrate pH 6.86 (FIRST – very important)
1. Turn the meter ON
2. Put probe into pH 6.86 solution
3. Wait until reading stops moving
4. Press and HOLD “MODE/CAL” for ~5–6 seconds
5. Screen will blink, then lock at 6.86
6. Release the button
✅ First calibration done
⸻
Step 4: Calibrate pH 4.01
1. Rinse probe with distilled water
2. Place probe into pH 4.01 solution
3. Wait for stable reading
4. Press and HOLD “MODE/CAL” again
5. Meter will adjust and lock at 4.01
⸻
Step 5: (Optional) Calibrate pH 9.18
Only needed if you test soap, detergents, alkaline solutions
• Same steps as above using pH 9.18
⸻
Step 6: Finish & store
• Rinse probe
• Add 2–3 drops of the protective/storage solution into the cap
• Close cap tightly
❌ Never store dry
❌ Never store in distilled water
⸻
Important tips (very important for accuracy)
• Always start with 6.86
• Recalibrate:
• Weekly (cosmetics / lab use)
• Monthly (light use)
• If readings jump or won’t stabilize → recalibrate